Quality Standards and Classification
Anti-electroshock degree:
Classle quipment (internal power supply)
Protection degree of fluid intake: IPX1
Application Range
• Adult children
Display
• 2.8inch display +LED(Red/Green)
Environment
working environment conditions
Ambient temperature range: +5°C~40 °C;
Relative humidity range: 15% ~ 80%, no condensation
Atmospheric pressure range: 60 ~ 107kPa;
Shipping and storage conditions
Ambient temperature: -20°C~+55°C
Atmospheric pressure: 16kPa ~ 107kPa
Humidity range: 10% ~ 95%
Electrical Specifications
Input: AC 100~240V AC,
Power frequency: 50/60Hz,
Maximum input power: 70VA
Fuse specification: 2A 16V
Working method: continuous work
Internal lithium battery: 2000 mAh
Dimension and Weight
Product Size 246.5*146.4*50.7mm
Product weight 950g
Optional accessories
Monitor trolley (optional)
Wall Mount (optional)
Features
• 2 hours of working time
Temperature
Range 0~50°C
Resolution 0.1°C
Accuracy +0.1°C (without probe error)
NIBP
Adult systolic blood pressure: 40 ~270mmHg
Diastolic blood pressure: 10~210mmHg
Average pressure: 20~ ~230mmHg
Children systolic blood pressure: 40 ~ 200mmHg
Diastolic blood pressure: 10 ~ 150mmHg
Average pressure: 20 ~ 165mmHg
Neonatal systolic blood pressure: 40 ~ 135mmHg
Diastolic blood pressure: 10~95mmHg
Average pressure: 20 ~ 110mmHg
Static pressure range: 0 ~ 300mmHg
Static pressure accuracy: 3mmHg
Blood pressure accuracy:
maximum average error 5mmHg;
maximum standard deviation 8mmHg
Overvoltage Protection:
Adult mode: 300 mmHg#10mmHg
Child mode : 240mmHg10mmHg
Neonatal mode: 150mmg#10mmHg
Sp02
Measuring range 0 ~ 100%
Resolution 1%
The accuracy is within the range of 70% to 100%, and the measure ment error should be $4%;In the range of 0% to 69%, the measure ment error is undefined
Pulse Rate
Measuring range 25 ~ 250bpm
Resolution 1bpm
Accuracy ≤3bpm
Refresh time 1s
Standard accessories
Sp02 sensor
TEMP probe
Cuff & Tube
Power cable
Ground cable
Lithium-ion battery
User Manual